ATF
  
  
  
  
  
  
  
  
  
  
  
    
  
  
    
  
    
      
      
        
      
      
        
          
            
                
                  
                    
                    
                      
                        
                          
                          
                          
                      
                    
                    
                      
                      
                      
                      
                        
                          
                      
                    
                  
                
              
          
            
          
        
        
        
      
    
  
  
  
  
    
    
  
    
        
 
      
  
Treatment
                        
                        - We provide this product in Silver and Gold, and the gold one has been heat-treated.
 - Silver Conventional NiTi, tough blade, great efficiency
 - Gold Heat-treated, increased cyclic fatigue resistance, enhanced flexibility
 
Operation Sequence
Create straight-line access to canal orifice.
- In the presence of a viscous chelator, passively scout the coronal 2/3 with 10# and 15# hand files.
 - Gently work these instruments until a smooth, reproducible glide path is confirmed.
 - In the presence of NaOCl, “float” the V1 into the canal and passively “follow” the glide path.
 - Before light resistance is encountered, laterally “brush” and cut dentin on the outstroke to improve straight line access and apical progression.
 - Continue shaping with the V1 as described until the depth of the #15 hand Fileis reached.
 - Use the V2, exactly as described for the V1, until the depth of the #15 hand Fileis reached.
 - In the presence of NaOCl or a viscous chelator, scout the apical 1/3 with #10 and #15 hand files and gently work them until they are loose at length.
 - Establish working length, confirm patency and verify the presence of a smooth, reproducible glide path in the apical 1/3.
 - Use the V1, as described, until working length is reached.
 - Use the V2, as described, until working length is reached.
 - Reconfirm working length, especially in more curved canals.
 - Use the T1 in a non-brushing action until working length is reached.
 - Gauge the foramen with a #20 hand file. If this instrument is snug at length, the canal is shaped and ready to obturate.
 - If the #20 hand File is loose at length, proceed to the T2 and,
 - when necessary, the T3, T4 and T5 gauging after each finishing File with corresponding hand file.
 
Basic Information
- Material: NiTi
 - Sterilization: ≤126°C
 - Package: 6 pcs per box/blister Assorted/single size
 - 1800 packings per cartoon 48cm*40cm*35cm
 - Color (Optional): Silver, Gold
 - Length (Optional): 21mm/25mm/31mm
 - Standard: CE/ISO/FDA/FSC
 

Treatment
                        - We provide this product in Silver and Gold, and the gold one has been heat-treated.
 - Silver Conventional NiTi, tough blade, great efficiency
 - Gold Heat-treated, increased cyclic fatigue resistance, enhanced flexibility
 
Operation Sequence
Create straight-line access to canal orifice.
- In the presence of a viscous chelator, passively scout the coronal 2/3 with 10# and 15# hand files.
 - Gently work these instruments until a smooth, reproducible glide path is confirmed.
 - In the presence of NaOCl, “float” the V1 into the canal and passively “follow” the glide path.
 - Before light resistance is encountered, laterally “brush” and cut dentin on the outstroke to improve straight line access and apical progression.
 - Continue shaping with the V1 as described until the depth of the #15 hand Fileis reached.
 - Use the V2, exactly as described for the V1, until the depth of the #15 hand Fileis reached.
 - In the presence of NaOCl or a viscous chelator, scout the apical 1/3 with #10 and #15 hand files and gently work them until they are loose at length.
 - Establish working length, confirm patency and verify the presence of a smooth, reproducible glide path in the apical 1/3.
 - Use the V1, as described, until working length is reached.
 - Use the V2, as described, until working length is reached.
 - Reconfirm working length, especially in more curved canals.
 - Use the T1 in a non-brushing action until working length is reached.
 - Gauge the foramen with a #20 hand file. If this instrument is snug at length, the canal is shaped and ready to obturate.
 - If the #20 hand File is loose at length, proceed to the T2 and,
 - when necessary, the T3, T4 and T5 gauging after each finishing File with corresponding hand file.
 
Basic Information
- Material: NiTi
 - Sterilization: ≤126°C
 - Package: 6 pcs per box/blister Assorted/single size
 - 1800 packings per cartoon 48cm*40cm*35cm
 - Color (Optional): Silver, Gold
 - Length (Optional): 21mm/25mm/31mm
 - Standard: CE/ISO/FDA/FSC
 